OSP-0394
Active
GE Healthcare
CT Systems
- OSP Number
- OSP-0394
- Manufacturer
- GE Healthcare
- Sub-Category
- CT Systems
- Category
- 8. CT Systems
- Expiration
- 3/17/2032
- HCAI Document
- View on HCAI
Seismic Parameters (per HCAI certificate)
Maximum design parameters under which this OSP was certified per ASCE 7 §13. Verify against the original HCAI document before specifying.
- SDS @ z/h = 1.0 (grade)
- 2.00 g
- SDS @ z/h = 0 (roof)
- 2.50 g
- SDS, max (g)
- 2.50
- z/h certified
- 1.5
- Raw SDS string (from certificate)
- 2.00 @ z/h = 1; 2.5 @ z/h = 0